6. When the correct Day displays, push the ON-OFF/PAGE
button once to advance to the Year.
7. Press and hold the CHANGE/RESET button to advance
the Year.
8. Once the correct year displays, push the ON-OFF/PAGE
button to leave the Date Set mode.
Accessing the Instrument Setup Mode
(FIGURES 4-2 and 4-3)
The Instrument Setup mode allows the user to change internal
values such as:
Default calibration gas values for autocalibration
Operating beep
Alarm set-points for exposure, STEL and TWA
Calibrating tolerance.
To Access the Instrument Set-up Mode:
1. While turning the instrument ON by using the ON-OFF/PAGE
button, push and hold the CHANGE/RESET button.
The Change flag flashes.
2. To enter the Set-up mode, push the ON-OFF/PAGE button.
Pushing the Change/Reset button returns the instrument to
the Measure mode.
The Change flag turns ON solid and stays ON as long as
the instrument is in the Set-up mode.
Chapter 4, Setting up the Multigas Detector Operating Manual
4-6