The LEL Banner turns ON solid.
•
The CAL flag turns ON.
5. Push the CHANGE/RESET button to increment the
autocalibration value.
6. Push the ON-OFF/PAGE button to accept the value.
OXYGEN SENSOR SET-UP
The Oxygen Sensor Set-up allows changes to be made to the:
•
Oxygen sensor high alarm (oxygen enrichment)
•
Oxygen sensor Low alarm (oxygen depletion).
•
The Oxygen banner flashes.
1. Push the ON-OFF/PAGE button to change the oxygen settings.
2. Push the CHANGE/RESET button to skip forward to the
CO Set-up.
•
The Oxygen banner turns ON solid.
•
The MIN (low or deficiency alarm) flag turns ON.
3. Push the CHANGE/RESET button to increment the
MIN alarm value.
4. Push the ON-OFF/PAGE button to accept the value.
•
The Oxygen banner turns ON solid
•
The PEAK (high or enrichment alarm) flag turns ON.
5. Push the CHANGE/RESET button to increment the PEAK
alarm value.
6. Push the ON-OFF/PAGE button to accept the value.
CARBON MONOXIDE (CO) SENSOR SET-UP
The CO Sensor Set-up allows changes to be made to the:
•
CO sensor exposure alarm (PEAK)
•
STEL alarm
•
TWA alarm and autocalibration value.
•
The CO banner flashes.
1. Push the ON-OFF/PAGE button to make changes to the
CO settings.