193
hyperion X5 3D CEPH/CEPH Ready - TECHNICAL MANUAL
EN
CIRCUIT BOARDS
DL17 Green 3.3V 3.3V CPU tension present, normally on
DL18 Green 24V 24V tension present, normally on
DL19 Green SYNC Flashing CBCT during frame acquisition
DL20 Green ENB
Inverter enabled signal from LOGIC board, normally on. O󰀨 during 
standby or in error condition.
DL21 ACK Not used
DL23 Green RDYL
External  “ready”  lamp  relay  status.  Normally  o󰀨,  on  when  unit  is 
“ready”.
DL24 Yellow XRYL
External “x-ray on” lamp relay status. Normally o󰀨, on during x-ray 
emission.
DL25 Green BUT Xray button pressed. On when x-ray button is pressed.
DL26 Yellow ILK
Interlock status. Normally o󰀨. On if door is open or interlock is not 
connected.
DL27 Red EMG
External  emergency  button  status.  Normally  o󰀨.  On  if  External 
emergency button status is pressed or not connected.
DL28 Yellow FIL X-ray lament on. On during emission.
DL29 Yellow INV Inverter on. On during emission
DL30 Yellow DRV Driver enabled, normally on. O󰀨 during standby or in error condition.
DL31 Green 48V 48V tension ON. O󰀨 during standby or in error condition.
DL32 Yellow ON
Mains  power  supply  to  power  ampier  relay  status.  O󰀨  during 
standby or in error condition.
DL33 Yellow PC
Preload  resistance  by-pass  relay.  O󰀨  during  standby  or  in  error 
condition.
DL35 Red EMG
Integrated emergency button status. Normally o󰀨. On if integrated 
emergency button status is pressed or not connected.
DL36 Red MT
Column engine overload. Normally o󰀨. On when thermal sensor is 
active or not connected.