Specifications
132 EasyOne Pro™ / EasyOne Pro™ LAB
Predicted val-
ues (spirome-
try)
NHANES III, Knudson_83, Knudson_76, Crapo, Mor-
ris, Hsu, Dockery (Harvard), Polgar, Gutierrez
(Canada), Eigen, Pereira, Pereira 2006/2008, Pérez-
Padilla (PLATINO), Pérez Padilla (Mexico), Pérez
Padilla (Mexico, Ped.), Chile 2010, 1997, ERS (ECCS/
EGKS), Zapletal 1977, Zapletal 2003, Stanojevic 2009
(GLI), Quanjer 2012 (GLI), Rosenthal, Austria 1988,
Austria 1994, Sapaldia (Swiss), Roca (Spain, SEPAR),
Vilozni 2005, Falaschetti, Klement (Russia), Heden-
ström (Sweden), Gulsvik, (Norway), Berglund, Birath
(Sweden), Langhammer (Norway), Finnish, Nystad,
Hibbert, Gore, Crockett, Ethiopia, Asia 1-4, JRS 2001,
Fukuda Standard (see "Predicted Values" on
page 139)
Predicted val-
ues (lung
capacity and
DLCO)
Ayers, Burrows, Crapo, Goldman & Becklake, Knud-
son, McGrath & Thompson, Miller, Gutierrez (Canada),
Klement (Russia), NHANES, Polgar, ERS, Zapletal,
Roca, Hedenström, Gulsvik, Pereira 2008, Thompson
(see "Predicted Values" on page 139)
Operating
conditions
Temperature 5 to 40 °C (41 to 104 °F)
LAB: 10 to 40 °C
(50 to 104 °F)
Relative humidity 15% to 95%,
no condensation
LAB: 30% to 75%,
no condensation
Atmospheric pressure 700 to 1060 hPa
Storage/
transport
Temperature -20 to 50 °C (-4 to +122 °F)
Relative humidity 5% to 95%, no
condensation
Atmospheric pressure 500 to 1060 hPa
Certifications CE Declaration of conformity
Device classi-
fication
Protection class I
Type BF applied part
Furthermore the device is not intended to be used in
the presence of oxygen rich environments. However,
where the device contains oxygen rich environments,
it complies with applicable requirements for fire pre-
vention.
Operating
voltage range
100 to 240 V, 50 to 60 Hz