EasyManuals Logo

NDD easyone pro User Manual

NDD easyone pro
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #84 background imageLoading...
Page #84 background image
Quality Messages and Quality Grades
84 EasyOne Pro™ / EasyOne Pro™ LAB
5 Quality Messages and Quality Grades
5.1 FVC/FVL Quality Messages and Quality Grades
Some Basic Facts
End-of-Test criteria, quality criteria and quality grading are based upon the
published standards [1], [4], [11], [15] ("Literature" on page 182).
Quality grading is based on [4], [11], [15] ("Literature" on page 182).
Additional remarks
The main articles [2] and [3] do not numerically define the minimum
expiratory peak flow time (PEFT) that is required for an acceptable test. In
this case, 160 ms is used.
The end-of-test criteria for FVC tests are as follows: A test ends when the
volume change during the last 2 seconds is <45 ml, or an inspiratory
volume >150 ml is detected. This end-of-test criterion slightly differs from
the published criterion in [3] which is 25 ml in the last second. Reason:
When the 25 ml in 1 second criterion is applied to the 24 waveforms also
defined in [1] and [2], several curves will end too early, and these tests
might fail. We therefore decided to slightly change the criterion to 45 ml
in 2 seconds.
In the following tables the quality messages and the quality criteria are
detailed.
The end-of-test criterion for an FVC test is as follows: A test ends when the
volume change during the last 2 seconds is <45 ml, or an inspiratory volume
>150 ml is detected. When this end-of-test criterion is met, the following
quality messages are checked:

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the NDD easyone pro and is the answer not in the manual?

NDD easyone pro Specifications

General IconGeneral
BrandNDD
Modeleasyone pro
CategoryMedical Equipment
LanguageEnglish

Related product manuals