Start-up and Initial Preparation
20 EasyOne Pro™ / EasyOne Pro™ LAB
2 Start-up and Initial Preparation
2.1 Unpacking, Environmental Conditions
Transport Damage
Upon arrival, immediately check that the shipment is complete and undam-
aged. If you have a complaint, promptly notify the shipping agent and your
local ndd dealer.
Check each of the following items:
the housing
the sensor and its connection cable
the valve unit including the overpressure valves and the one-way
valve
the gas supply tube
the power cord
the gas cylinder fittings.
Cleaning Before Initial Use
EasyOne Pro™ does not require cleaning before its first use.
The gas supply tube and the valve unit are cleaned before shipping, but not
disinfected. These parts can be thoroughly cleaned as described in chapter
"Hygiene, Cleaning, Maintenance, Disposal" on page 100.
Ambient Conditions
Excessive amounts of dust, lint, and miscellaneous clutter around the
instrument could result in malfunctions due to internal tubing blockages,
overheating of components, clogged ventilation ports, etc.
EasyOne Pro™ can be operated under the following ambient conditions:
temperature +5 to +40 °C (41 to 104 °F)
(LAB: 10 to 40 °C (50 to 104 °F))
relative humidity 15 to 95%, no condensation
(LAB: 30 to 75%, no condensation)
atmospheric pressure 700 to 1060 hPa.
Note
Also observe the information given in section
"Electromagnetic Compatibility (EMC)" on page 154.