EasyOne Pro™ / EasyOne Pro™ LAB 131
Specifications
10 Specifications
Dimensions 270 x 335 x 270 mm
3
(h x w x d)
Weight [kg] < 8 kg, LAB < 9 kg
Flow/volume
measurement
Accuracy, volume: ±2% or 0.050 l
Accuracy, flow: ±2% or 0.020 l/s
(except PEF)
Accuracy, PEF: ±5% or 0.200 l
Accuracy, MVV: ±5% or 5 l/min
Resolution, volume >1 ml
Resolution, flow >4 ml/s
Measuring range, volume ±12 l
Measuring range, flow ±16 l/s
Resistance <1.5 cm H
2
O/l/s at 12 l/s
Molar mass
analyzer
Type ultrasonic transit time
Measuring range 9 to 50 g/mol
Accuracy <0.02 g/mol
Resolution 0.01 g/mol
Carbon mon-
oxide analyzer
Type NDIR
Measuring range 0 to 0.35%
Accuracy ±0.001%
Resolution 0.0001%
CO2 analyzer
(LAB only)
Type nondispersive infrared
Measuring range 0 to 15%
Resolution 0.005%
Accuracy from 0 to 5%:±0.05%
from 5 to 10%:±0.20%
>10%: not specified
Mouth pres-
sure trans-
ducer
Measuring range ±100 mbar
Accuracy ±5%