Appendix
144 EasyOne Pro™ / EasyOne Pro™ LAB
View of tests with EasyOne Pro™
View test via GDT record type 6311 with patient name, ID and date and
time of test
Record type 6311 is supported. In the transmission of test results using GDT
record type 6310, field IDs 8432 (date) and 8439 (time) are always
included:
8432ddmmyyyy (day, month, year)
8439hhmmss (hours, minutes, seconds)
In case of a recall via GDT record type 6311, the test can be identified by
patient name, ID, test date and test time.
Example "forced expiratory spirometry"
Here the example transmission GDT Export with a forced loop test (FVL). The
data marked grey will be exported only when the "Export single measure-
ment value" check box is selected.
01380006310
014810001385
0128315EDV
0128316NDD
014921802.00
01230006-0
0183101Appeldorn
0173102Heinrich
017310323101945
01031101
0123622178
011362366
0158402LUFU02
017620024042008
017843224042008
017843918210004
0366228Your FEV1 / Predicted: 103%
0096228
0436228----------------------------------
0436228 Pre
0436228Parameter Pred Best %Pred
0436228----------------------------------
0436228FVC L 4.72 4.03 85
0436228FEV1 L 3.55 3.67 103
0436228FEV1/FVC 0.75 0.91 121