WARNING: Disconnect the N-395 and Nellcor sensor from
the patient during magnetic resonance imaging (MRI)
scanning. Induced current could potentially cause burns.
The N-395 may affect the
MRI
image and the MRI unit may
affect the accuracy of
oximeter measurements.
WAmING: To ensure accurate performance and prevent
device failure, do not subject the N-395 to extreme moisture,
such as direct exposure to rain. Such exposure may cause
inaccurate performance or device failure.
I
WARNING: Do not use an N-395 monitor, sensor, cables, or
1
I
connectors that appear to be damaged.
1
WARNING:
The N-395 is not defibrillator-proof. However,
it may remain attached to the patient during defibrillation or
while an electrosurgical
unit is in use, but the readings may
be inaccurate during use and shortly thereafter.
Connecting the
N-395
to
AC
Power
The N-395 operates on AC power when the hospital-grade power
cord is connected to both
the inonitor and an AC power source
(wall outlet).
The supply voltage selector switch allows
connection of the
monitor to AC power ranging from
100 volts AC to 240 volts
AC.
The switch has two positions: one for 100
-
120 volts AC
("1
15"),
and one for
200
-
240 volts AC ("230").
Ensttre that the
supply voltage selector switch on the
reap panel is set to the
proper voltage.
Operating
on
a
Discharged
Battes~gl
The N-395 will not operate when its internal battery is
coinpletely discharged,
even when the inonitor is connected to
AC power. Instead, the en-or code
"EEE
04" will be displayed.
This feature prevents the accidental use of the monitor with a
dead battery. The monitor is only capable of indicating a loss of
AC
power if its internal battery is functional.