Datex-Bhmeda Communications
The Datex-Ohmeda
inonitor AS13 must be configured for
cormnuilications with the Nellcor N-200 monitor in order to
cormnuilicate with the W-395 monitor. Refer to the AS13
operator's manual for instructions on configuring the AS13
monitor.
The N-395 sends
SpOz, pulse rate, and alarm status data to the
Datex
AS13 monitor.
The RS-232
hardwire interface cable has a DB-15 connector
for the N-395
and the applicable connector for the Datex
monitor. Nellcor cable part number 902255 is recommended for
this interface.
Note:
Spare parts for the N-395 are listed
on the Internet,
confirm the cable part number before ordering a cable.
The Intenlet address is:
http:llmallinckrodt.codrespiratoryiresp/§erv
-
SupplApa
rtweb/inaii~/PartAcceMei~u.htinl.
Corrupt data will be indicated by a Commui~ications Error
displayed on the Datex monitor. The Datex
moilitor will detect
compt data in less than 11 seconds.
When the N-395 is in the Datex mode of operation the interface
baud rate must be set to 2,400 bits per second. Pressing the
SETUP
softkey, then the NEXT softkey, and then the
COMM
softkey to select BAUD. Use the ADJUST UP or ADJUST
DOWN
buttons to select the correct baud rate.
WARNING:
Do not silence the
N-395
audible alarm or
decrease its volume if patient safety could be compromised.
The Datex-Ohmeda monitor does not indicate audible or visual
alaims. The N-395 monitor must be able to sound an audible
alarn in order to maiiltaiii patient safety.
LANG
The LANG softkey provides selection of the language displayed
on the screen.