Press the NEXT softkey to access the LANG softkey. After
pressing the
LANG softkey, use the ADJUST UPIDOWN
Buttons to select English, French, German, Italian, Spanish,
Dutch, or Portuguese.
If the language is changed and EXIT is pressed (or a 10-second
tiineout occurs),
the monitor begins displaying data in the
selected language.
MCALL
The NCALL softkey provides the capability of setting the alarm
voltage at a normally high (NORM
+)
or normally low
(NORM
-).
Refer to the "Nurse Call" heading of
Data Port
Protocol
for a more thorough explanation of these settings. The
nurse call feature is also discussed later in this section.
ANALOG
The ANALOG softkey provides the capability to produce
variable calibrating voltages to calibrate
instruinents such as a
chart recorder. Refer to the "Analog
Outputs" heading of
Data
Port Protocol
for a more thorough explanation of these settings.
The LIGHT
softkey turns the backlight on or off. When the
backlight is off, pressing any softkey turns the backlight on.
Also, the CONTRAST
and ALARM SILENCE keys will
turn
the backlight on. Any alarin will turn the backlight on. Turning
the backlight off conserves battery power.
Default
Settings
The N-395 is shipped with factory default settings (Table
6
and
Table
7).
Authorized technical personnel using the procedures
described in the N-395 service manual can change default
settings.
Note: Factory default settings are
constaiits that cannot be
changed without
re-comnpiling software.