32
Fn
Filter ( same as averaging ) Slow, Medium, Fast
Function: This command is used to set how many measurements are
averaged for the displayed reading. When the slow speed
filter is selected, the 1830-C displays the average of the last 16
measurements. When using the medium filter, the 1830-C
displays the average of the last 4 measurements. The fast
filter does no averaging.
Syntax: Fn or fn
Parameter: n = 1 use the slow filter to average the last 16 measurements
n = 2 use the medium filter to average the last 4 measurements
n = 3 no averaging is performed
Type: integer
Related
Commands: F? - Filter status query
F?
Filter (same as averaging) status query
Function: This query informs the user about the current settings for
the filter parameter. The filter setting dictates how many
measurements are averaged for the displayed reading. In the
slow mode, the reading is the average of the last 16 measure-
ments, while in the medium mode the reading is the average
of the last 4 measurements. No averaging is done in the fast
mode.
Syntax: F? or f?
Parameter: none
Returns: 1 The readings are being averaged in slow mode
2 The readings are being averaged in medium mode
3 The readings are not being averaged
Related
Commands: Fn - Set how many measurements are averaged for the
displayed reading