43
Zn
Zero on/off
Function: Turn the zero function on/off. Zero function is used for
subtracting any background power levels from future mea-
surements. When it is activated, the 1830-C uses the next
power reading as it’s background value, and it subtracts this
value from any future power readings.
Syntax: Zn or zn
Parameter: n = 0 zero function is off
n = 1 zero function is on
Type: integer
Related
Commands: Z? - Zero function query
Z?
Zero function query
Function: This query informs the user whether 1830-C’s zero function is
on/off. Zero function is used for subtracting the background
power levels from any future measurements.
Syntax: Z? or z?
Parameter: none
Description:
Returns: 0 zero function is off.
1 zero function is on.
Related