Chapter 7 Calibration  53 
 
 
7.2.3  Remote Operation Current Source Calibration 
a.  Connect the shunt to the laser outputs. Connect a calibrated 
DMM across the shunt’s terminals to measure the voltage. 
Calculate the current in the following steps by using Ohm's Law: 
I = V / R 
 
  where V is the measured voltage across the resistor, and R is the 
measured load resistance. 
 
b.  Select the channel via the 
LAS:CHAN
 command. Set the 
current limit to one-half scale plus 100 mA via the 
LAS:LIM:LDI
 command, output bandwidth as desired via the 
LAS:MODE:I
, 
LAS:MODE:ICW
, or 
LAS:MODE:ILBW
 
command, and current set point to one-half scale via the 
LAS:LDI
 command. 
 
c. Enter the 
LAS:OUTPUT ON
 command to turn the laser output 
on. 
 
d.  Enter the laser LDI calibration mode by issuing the 
LAS:CAL:LDI
 command. 
 
e.  Input the actual (measured) laser output current (as an <nrf 
value>) via the 
LAS:LDI <nrf value>
 command. 
 
  The 5600 will be ready to receive the current value when, after a 
LAS:CAL:LDI?
 query is sent, the response from the 5600 is 
"1".  
 
f.  Once the actual current value is entered via the 
LAS:LDI
 
command, the 5600 will apply a new current equal to 
approximately one-fourth (
¼
) the previous set point. 
 
  The 5600 will be ready to receive the second current value 
when, after a 
LAS:CAL:LDI?
 query is sent, the response from 
the 5600 is "1".  
 
g.  Input the second actual (measured) laser output current as in 
Step e. 
 