EasyManua.ls Logo

Nordson Ultimus V - Set Trigger Value Command; Set the Real Time Clock Command

Nordson Ultimus V
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.6.2.20 Set Trigger Value Command
This command downloads the 5-digit trigger value into the current memory location.
Client command and data: EQ--Tttttt
ttttt: Trigger Value. 00001–99999
EXAMPLE: If the trigger value is 1000, the text packet it: (STX) + 0A + EQ-- + T01000 + 74 + (ETX)
Length of this text packet: Decimal 10 or hexadecimal 0A
The above text packet in hexadecimal format:
STX 0 A E Q space space T 0 1 0 0 0 7
0x02 0x30 0x41 0x45 0x51 0x20 0x20 0x54 0x30 0x31 0x30 0x30 0x30 0x37
4 ETX
0x34 0x03
2.6.2.21 Set the Real Time Clock Command
This command sets the time for the real time clock on the dispenser.
Client command and data: EB--HhhMmmAMa
hh: Hours. 0–23 for 24 hour format, 1–12 for 12 hour format
mm: Minutes. 0–59
a: Hour format. 0 = AM, 1 = PM, 2 = 24 hour format
EXAMPLE: To set the time as 14:05 and the hour format to 24-Hour, the text packet is:
(STX) + 0D + EB-- + H14M05AM2 + A6 + (ETX)
Length of this text packet: Decimal 13 or hexadecimal 0D
The above text packet in hexadecimal format:
STX 0 D E B space space H 1 4 M 0 5 A
0x02 0x30 0x44 0x45 0x42 0x20 0x20 0x48 0x31 0x34 0x4D 0x30 0x35 0x41
M 2 A 6 ETX
0x4D 0x32 0x41 0x36 0x03
Appendix B — RS‑232 Protocol (continued)
Ultimus V High Precision Fluid Dispenser
59www.nordsonefd.com info@nordsonefd.com 800-556-3484 Sales and service of Nordson EFD dispensing systems are available worldwide.

Table of Contents