2.6.3.12 Operator Lockout Read Command (continued)
NOTE: “D0” is the letter “D” followed by the digit “0.”
Return Format: D0DTtDPpDVvMmDCcDMdAIaARuALbMMePUfVUgLAhCLjCOkAMn
t: Lockout dispense time: “1”=lockout, “0”=enabled (DT)
p: Lockout dispense pressure (DP)
v: Lockout dispense vacuum (DV)
m: Lockout memory cell selection (M)
c: Lockout deposit counter selection (DC)
d: Lockout dispense mode change (DM)
a: Lockout Auto Increment Mode selection (AI)
u: Lockout Auto Increment Reset (AR)
b: Lockout Alarms Reset (AL)
e: Lockout Main Menu selection (MM)
f: Lockout Pressure Unit Menu selection (PU)
g: Lockout Vacuum Unit Menu selection (VU)
h: Lockout Language Menu selection (LA)
j: Lockout Set Clock / Date Menu selection (CL)
k: Lockout Set Communications Menu selection (CO)
n: Lockout Alarm Options Menu selection (AM)
EXAMPLE: If the Dispense Time (DT), Dispense Pressure (DP), and Dispense Vacuum (DV) are locked out, the text
packet received from the dispenser is:
(STX) + 31 + D0DT1DP1DV1M0DC0DM0AI0AR0AL0MM0PU0VU0LA0CL0CO0AM0 + 2A + (ETX)
Length of the response text packet: Decimal 49 or hexadecimal 31
The above text packet in hexadecimal format:
STX 3 1 D 0 D T 1 D P 1 D V 1
0x02 0x33 0x31 0x44 0x30 0x44 0x54 0x31 0x44 0x50 0x31 0x44 0x56 0x31
M 0 D C 0 D M 0 A I 0 A R 0
0x4D 0x30 0x44 0x43 0x30 0x44 0x4D 0x30 0x41 0x49 0x30 0x41 0x52 0x30
A L 0 M M 0 P U 0 V U 0 L A
0x41 0x4C 0x30 0x4D 0x4D 0x30 0x50 0x55 0x30 0x56 0x55 0x30 0x4C 0x41
0 C L 0 C O 0 A M 0 2 A ETX
0x30 0x43 0x4C 0x30 0x43 0x4F 0x30 0x41 0x4D 0x30 0x32 0x41 0x03
Appendix B — RS‑232 Protocol (continued)
Ultimus V High Precision Fluid Dispenser