EasyManua.ls Logo

NSK Surgic Pro2 - Page 20

NSK Surgic Pro2
55 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
19
Operation Procedure
English
Speed calibration
1
Remove the calibration bur from the calibration load�
2
Press (CALIBRATION) with the bur not contacting
anywhere�
3
A beep sounds and the calibration bur automatically
rotates for a few seconds at low speed�
4
Calibration is complete when a beep sounds and
“PASS” appears on the display� Proceed to next step�
Speed calibration
1
Press (CALIBRATION) with the bur not contacting
anywhere�
2
A beep sounds and the calibration bur automatically
rotates for a few seconds at high speed�
3
Calibration is complete when a beep sounds and
“SUCCESS” appears on the display�
4
The LCD display will return to the normal screen�
Calibration fail example
● If a problem occurs such as the gear being extremely worn, correct data cannot be obtained at calibration and
“FAIL” is displayed� In this case, please request repair�
● Even if a malfunction does not occur, “FAIL” is displayed if the bur comes in contact with something during no-
load calibration or if you forget to apply a load at load calibration�

Table of Contents