13
6 Optional Parts
1. T1-MODULE
2. T2-MODULE
6.1 T1-MODULE
The T1-Module (see Figure 14) is a calibrated temperature probe that can be read and displayed by LEO.
The T1-MODULE has an external diameter of 0.13 mm and it’s suitable for temperature measurements of benchtop
and hightop incubators where the temperature monitoring port has a diameter smaller than 2mm. T1-MODULE
stores calibration data in its own local memory
Tip Calibration data is available in the Calibration page of LEO when T1-MODULE is connected to LEO
(see paragraph 12).
Figure 14. T1-MODULE
Note Okolab recommends that the T1-MODULE should be calibrated after one year of use.
6.2 T2-MODULE
The T2-MODULE (see Figure 15) is a calibrated temperature probe that can be read and displayed by LEO.
The T2-MODULE has an external diameter of 2 mm and it’s suitable for temperature measurements of benchtop
and hightop incubators where the temperature monitoring port has a diameter bigger than 2 mm. T2-MODULE
stores calibration data in its own local memory.
Tip Calibration data is available in the Calibration page of LEO when T2-MODULE is connected to LEO
(see paragraph 12).
Figure 15. T2-MODULE
Note Okolab recommends that the T2-MODULE should be calibrated after one year of use.