Chapter 7
54
7.10.1 Resetting Measurement Parameters
The Measurement Reset function resets the measurement parameters to their default
values.
The parameters that are reset and their reset values are as follow:
Material Velocity (0.5740 mm/µs [0.2260 in/µs])
Transducer Zero
Measure Option (Standard)
Unit (Inches)
Resolution (Standard)
Min/Max (Off)
Hold/Blank (Blank)
•Measure Rate (Normal)
•CAL Lock (Off)
•Beeper (On)
Inactive Time (On)
•Radix (Period)
Backlight mode (Normal)
Alarm (Off). Default Values: Low 0.000; High 25.000
Diff (Off). Default Value 0.000
To reset measurement parameters
1. Press [2
nd
F] [] (SETUP) to display the setup tabs.
2. Use the [, ] keys to highlight the RESET tab.
3. Use the [] key to highlight MEASUREMENT RESET and press [ENTER].
4. Use the [, ] keys to highlight RESET or CANCEL and press [ENTER].
5. Press [MEAS] to return to measurement mode.
7.10.2 Resetting the Entire Instrument Parameters
The Master Reset function resets the entire instrument parameters to their default
values.