Olympus 27MG - User Manual

Olympus 27MG
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
DMTA-10043-01EN — Rev. C
July 2016
27MG
Ultrasonic Thickness Gage
Users Manual
This instruction manual contains essential information on how to use this Olympus product safely and effectively.
Before using this product, thoroughly review this instruction manual. Use the product as instructed.
Keep this instruction manual in a safe, accessible location.

Table of Contents

Other manuals for Olympus 27MG

Questions and Answers

  • D
    Devon YatesJul 27, 2025
    What to do if Olympus Measuring Instruments measurements cannot be made and the MEAS and LOS flags are on?
    • E
      espinozatammieJul 28, 2025
      If you're unable to take measurements and see the MEAS and LOS flags on your Olympus Measuring Instruments, ensure you're using enough couplant, especially on uneven surfaces. Test the transducer on a smooth, flat sample. Adjust the Gain or Material Sensitivity. If problems persist, try a more sensitive transducer for your thickness range. If issues continue, the gage or transducer might need repair or replacement by Olympus.
  • C
    Casey PriceAug 4, 2025
    What to do if Olympus 27MG Ultrasonic Thickness Gage turns off immediately or does not turn on?
    • M
      matthewayalaAug 4, 2025
      If your Olympus 27MG Ultrasonic Thickness Gage isn't turning on, try replacing the batteries. If it still fails to power up after replacing the batteries, it likely indicates an internal component issue that requires servicing.
  • K
    Kevin CarrollAug 12, 2025
    What to do if message Do?? will not go away when the [ZERO] key is pressed on Olympus 27MG?
    • W
      William MillerAug 12, 2025
      If the 'Do??' message persists when pressing the [ZERO] key on your Olympus Measuring Instruments, ensure an Olympus transducer is properly connected. If it is, the transducer might be faulty; try using a different one or a different cable. If the issue persists across multiple transducers, the pulser/receiver assembly of the gage likely has a problem.
  • S
    Stephanie WeissAug 18, 2025
    Why does Olympus 27MG PR TX display N/A?
    • U
      unewmanAug 18, 2025
      If your Olympus Measuring Instruments display 'N/A' for PR TX, it means either the cable is broken, or there is an issue with the transducer.

Summary

Important Information - Please Read Before Use

Intended Use

Defines the primary applications and purpose of the 27MG Ultrasonic Thickness Gage.

Safety Symbols

Explains the meaning of various safety-related symbols used on the instrument and in the manual.

Safety Signal Words

Details the DANGER, WARNING, and CAUTION signal words used to indicate hazard levels.

Warnings

Provides general warnings and precautions for safe operation and handling of the instrument.

Battery Precautions

Outlines essential safety guidelines and procedures for handling and disposing of batteries.

Warranty Information

Details the terms and conditions of the Olympus product warranty and limited liability.

Instrument Description

Product Description

Overview of the 27MG Ultrasonic Thickness Gage, its design, and capabilities for thickness measurement.

Instrument Hardware Components

Describes the physical parts and layout of the 27MG instrument, including display and keypad.

Keypad Functions

Explains the function of each key on the English and international keypads for operating the instrument.

27 MG Power Requirements

Batteries

Covers battery types, operating time, level indicator, storage, and replacement procedures for the 27MG.

Software User Interface Elements

Measurement Screen

Details the main display elements of the 27MG software, including thickness value and indicators.

Parameter Screens

Explains how to navigate and configure instrument settings through various parameter tabs and menus.

Initial Setup

Standard Calibration Measurement

Transducer Zero Compensation

Describes the procedure for calibrating transducer zero offset for accurate measurements.

Material Velocity Calibration

Explains how to calibrate the material sound velocity for accurate thickness readings.

Measurements

Additional 27 MG Gaging Features

Configuring the Measurement Setup

Details how to adjust measurement parameters like units, resolution, and Min/Max mode.

Configuring the System Setup

Explains how to configure system-level settings such as beeper, inactive time, and backlight.

Activating High;Low Alarms

Guides on setting high and low alarm thresholds for thickness measurements.

Activating Diff Mode

Describes how to enable and use the differential thickness measurement mode.

Specifications

Theory of Operation

Application Notes

Factors Affecting Performance and Accuracy

Discusses various factors that can influence the accuracy and performance of thickness measurements.

Transducer Selection

Provides guidance on selecting appropriate transducers for different measurement scenarios and materials.

High Temperature Measurements

Offers special considerations and procedures for performing measurements at elevated temperatures.

Maintenance and Troubleshooting

Routine Care and Maintenance

Outlines procedures for cleaning and maintaining the 27MG instrument to ensure longevity.

Error Messages

Lists common error messages and their potential causes or solutions.

Measurement Problems Diagnostic

Provides steps to diagnose and resolve issues when measurements cannot be obtained.

Self Diagnostics

Explains how to use the instrument's built-in diagnostic screens to identify hardware or software problems.

Gage Performance Tests

Details procedures for testing the functionality of the keypad and display of the 27MG.

Repair Service

Information on how to obtain repair services for the 27MG Ultrasonic Thickness Gage.

Appendix: Sound Velocities

Olympus 27MG Specifications

General IconGeneral
Operating Temperature14° F to 140° F (-10° C to 60° C)
Weight9.7 oz (275 g)
Resolution0.01 mm
Battery2 x AA batteries
Frequencies2.25MHz, 5MHz

Related product manuals