EasyManuals Logo

Omicron Bode 100 User Manual

Omicron Bode 100
144 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #135 background imageLoading...
Page #135 background image
135
Technical Data
12 Technical Data
12.1 Bode 100 Specifications
Table 12-1:
Bode 100 specifications
Characteristic Rating
Frequency range
(selectable by the
Bode Analyzer Suite)
10 Hz…40 MHz or
1Hz40MHz
(extended frequency range)
OUTPUT connector
Output impedance 50 Ω
Connector BNC
Wave form Sinusoidal signal
Output voltage 0.01…1 Vrms into 50 Ω load
–27 dBm…13 dBm
CH 1 INPUT, CH 2 INPUT connectors
Input impedance Low or high impedance selectable
Low impedance Input impedance 50 Ω
High impedance
Input impedance 1 MΩ ±2%
Input capacitance 40…55 pF
Connectors BNC
Receiver bandwidth 1 Hz, 3 Hz, 10 Hz, 30 Hz, 100 Hz,
300 Hz, 1 kHz, 3 kHz
Input attenuator 0 dB, 10 dB, 20 dB, 30 dB, 40 dB
Input sensitivity 100 mV full scale
for input attenuator 0 dB
Dynamic range > 100 dB at 10 Hz receiver bandwidth
Gain error < 0.1 dB (calibrated)
Phase error < 0.5º (calibrated)
USB interface
Connector Type B