EasyManua.ls Logo

Omron E5AR - MANUAL 2

Omron E5AR - MANUAL 2
343 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8.6 Bank setting level (l.bnk)
8-27
Setting data
Use this setting to set the SP (local SP) in each bank.
The SP of banks 0 to 7 can be set.
When an SP is changed in "PV/SP" in Operation level, the local SP
of the currently used bank is also changed.
Related information
"5.2 Control functions Banks" (P.5-9)
Related setting data
"PV/SP" (Operation level) (P.8-7)
Use this setting to store a PID Set No. in each bank.
The PID Set No. of banks 0 to 7 can be stored.
When the PID Set No. is set to "0", the PID set auto select function
automatically selects a PID set based on the PV or DV (deviation). If
you wish to specify a PID set, set the number of the PID set (1 to 8).
Related information
"5.2 Control functions Banks" (P.5-9)
"5.2 Control functions PID sets" (P.5-12)
Related setting data
"PID
proportional band", "PID Integral time", "PID derivative
time" (PID setting level) (P.8-31)
"PID
Automatic selection range upper limit" (PID setting level) (P.8-
32)
"PID set automatic selection data" (Expansion control setting level)
(P.8-74)
Bank
local SP (LSP)
(
: 0 to 7)
.lsp
CH
l.bnk
Function
Setting
Setting data Setting range Units
Default value
Local SP
SP lower limit to
SP upper limit
EU 0
Reference
Reference
Bank PID Set No.
(
: 0 to 7)
.pid
CH
l.bnk
Function
Setting
Setting data Setting range Units
Default value
PID Set No. 0 to 8 0
Reference
Reference

Table of Contents

Related product manuals