EasyManua.ls Logo

OncoTherm EHY-2000plus - Page 45

OncoTherm EHY-2000plus
86 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
plus
UM - EHY2000plus-R24
44
If the „Treatment power” is much less than the „Forwarded power” (the
Reflected power is more than 25% of the „Forwarded power”), use the „Re-
tune” button to tune the device to the patient. If the re-tuning is not successful,
please pause the treatment by the „Pause” button, and check the correct
placement of the bolus electrode. You can continue the paused treatment by
the „Start” button. To abort the running treatment you can use the „Stop”
button. On the appearing of the „Treatment stop” screen you can choose to
view the logged data of the treatment or get back to the Main menu.
During the treatment the treatment parameters can also be modified. To
change the treatment power there are two possibilities. To change the
treatment power to a certain value you should click on the “Set power” button
on the “Treatment management view” screen. On the appearing “Set