transport, inform the carrier and do not use the
device.
• For long-distance transport (eg dismantling, return for
repair, etc.), the device must be placed in its original
packaging or in special packaging. For further
information, contact your dealer or the OPTIKON 2000
S.p.A. technical service.
Assembly and installation
INFORMATION Risk of damaging the Medical Device!
• Make sure that the installation and use conditions of
the device comply with the surgical requirements:
– Low vibration
– Clean environment
– Assess of excessive mechanical loads
INFORMATION Risk of interferences and malfunctions
!
• This Medical Device requires special precautions
regarding EMC (Electomagnetic Compatibility), it must be
installed and commissioned according to the EMC
information and indications given in the accompanying
documents.
INFORMATION Risk of interferences and malfunctions
!
•
This Medical Device should not be used in proximity of
other devices. If it must be used close to other
equipment, it must be carefully observed to verify
normal operation in the actual configuration in which it
is used.
INFORMATION Risk of damaging the Medical Device!
The maximum height of the IV pole is 225 cm from the
ground.
• Do not install the R-EVO SMART systems in rooms with a
low ceiling.
INFORMATION Risk of damaging the Medical Device!
If the ventilation openings are closed and blocked, the
device may overheat.
• Install the R-EVO SMART so that the ventilation
openings are neither closed nor obstructed.
WARNING Risk of explosion or fire hazard!