The Medical Device is not suitable for use in hazardous
areas.
•
R-Evo Smart systems must not be used:
– in areas at risk of explosion.
– if inflammable anesthetics or volatile solvents,
such as alcohol, benzene or similar chemicals,
are present at a distance of less than 25 cm.
• Do not use or store the system in damp rooms. Do not
expose the system to water splashes, dripping water or
sprayed water.
• To ensure safe operation, do not install the Medical
Device in a location that may be exposed to heaters or
radiators, direct sunlight, or any other heat source at
excessive temperatures.
Operation
CAUTION Risk of injury to the patient or the user!
• Observe the maximum loads indicated for the following
components:
– Maximum load on the instrument tray, 1 kg.
– Maximum load of the bottle and irrigation solution, 0.5
kg.
CAUTION Risk of injury to the patient or the user!
Danger of overturning when crossing doorsteps.
• Push the medical device slowly and carefully using its
handles when crossing doorsteps up to 3 cm. If the
doorstep is higher than 3cm it must be moved by two
people.
INFORMATION Risk of minor injury to the patient!
A moving white dot logo at the bottom right of the
screen indicates that the device is working properly.
• If the dots around the logo stop moving, the
device is stopped, and you must stop using it.
• Before any use, carry out the installation, the surgical
configuration and the operating procedures described
here. If a malfunction occurs that is not corrected using
the "Correction of malfunctions" chapter, please put the