R-evo Smart E/S/CR Installation and use manual
Cod. 1X10XXEN 2020-09-14 Rev.E
9
The Medical Device is not suitable for use in hazardous
areas.
R-Evo Smart systems must not be used:
in areas at risk of explosion.
if inflammable anesthetics or volatile solvents,
such as alcohol, benzene or similar chemicals,
are present at a distance of less than 25 cm.
Do not use or store the system in damp rooms. Do not
expose the system to water splashes, dripping water or
sprayed water.
To ensure safe operation, do not install the Medical
Device in a location that may be exposed to heaters or
radiators, direct sunlight, or any other heat source at
excessive temperatures.
Operation
CAUTION Risk of injury to the patient or the user!
Observe the maximum loads indicated for the following
components:
Maximum load on the instrument tray, 1 kg.
Maximum load of the bottle and irrigation solution, 0.5
kg.
CAUTION Risk of injury to the patient or the user!
Danger of overturning when crossing doorsteps.
Push the medical device slowly and carefully using its
handles when crossing doorsteps up to 3 cm. If the
doorstep is higher than 3cm it must be moved by two
people.
INFORMATION Risk of minor injury to the patient!
A moving white dot logo at the bottom right of the
screen indicates that the device is working properly.
If the dots around the logo stop moving, the
device is stopped, and you must stop using it.
Before any use, carry out the installation, the surgical
configuration and the operating procedures described
here. If a malfunction occurs that is not corrected using
the "Correction of malfunctions" chapter, please put the