4. INSTALLATION AND OPERATION
Before the first start
CAUTION Risk of injury to the patient's eye!
The Medical Device must be installed by the technical
service of OPTIKON 2000 S.p.A. or by an expert
authorized by it. Ensure that the following
requirements are met for the entire duration of
operation:
•
The connection parts are inserted correctly. The
connection screws are tightened.
•
All cables and plugs are in perfect condition, e.g. show
no signs of wear, kinks or other damage.
•
The rated voltage of the instrument corresponds to the
voltage of the network line at the place of installation.
•
The plug can only be connected to an outlet
equipped with a earth protection conductor in
perfect condition.
•
The device is connected to the power cord supplied by
OPTIKON 2000 S.p.A.
Before each use
CAUTION Risk of injury to the patient's eye!
• Ensure that all specified "Operation Requirements" are
observed.
• Reconnect the covers or plugs removed on the front.
Close all existing openings with their plugs.
• Make a note of all the symbols and signs posted on the
Medical Device.
• The ventilation openings must not be closed or covered.
• Check that the R-EVO SMART is located in a place
where there is sufficient space for manoeuvring in
order to avoid damaging the cables, to guarantee total
freedom of movement of the Medical Device and that
it has easy access to the power switch to disconnect
the appliance from the mains.
• Connect the equipotential connector on the back of