R-evo Smart E/S/CR Installation and use manual
Cod. 1X10XXEN 2020-09-14 Rev.E
81
WARNING Risk of injury to the patient's eye!
To reduce the risk of damage to the retina, the end of
the fibre optic light probe must not be too close to the
retina.
INFORMATION Duration of Surgical Operation
If the light is on and its connector is not engaged
(optical fibre not connected) The R-EVO SMART
automatically decreases the light output so as not to
disturb the surgeon.
Select the source you want to adjust and turn it on/off
using the buttons <on/off>,
Adjust the illumination intensity of the selected lamp
using the <+ -> keys or the cursor.
The surgeon can choose one of four filters designed to
emphasize the presence of membranes.
Moreover, the yellow filters protect the patient's eye from
unnecessary exposure to blue radiation and therefore allow
long-lasting treatments.
4.5.6 Tamponade
WARNING Risk of injury to the patient's eye!
To avoid possible infection of the patient's eye, only
use the original OPTIKON 2000 S.p.A. injection tube
equipped with an air sterilisation filter for air
tamponade.
CAUTION Risk of injury to the patient!
Automatic monitoring of intraocular pressure is
calibrated for the exclusive use of accessories
manufactured and distributed by OPTIKON 2000 S.p.A.
(membrane filters, tubings, scleral guides).
WARNING Risk of injury to the patient's eye!