R-evo Smart E/S/CR Installation and use manual
Cod. 1X10XXEN 2020-09-14 Rev.E
82
The silicone stop valve is checked during power on
test. Make sure there is no hissing when turning the
device on.
CAUTION Risk of injury to the patient's eye!
During the injection of silicone oil, the intraocular
pressure (IOP) is not controlled by the R-Evo Smart
CR, so it is the Surgeon who must take care of the
same. It is important to check that the silicone oil
flow is consistent with the injection pressure set
before inserting the cannula into the eye.
CAUTION Risk of injury to the patient's eye!
Release the pedal to stop the silicone oil injection. In an
emergency, close the injection stopcock or remove the
silicone oil injection tube from the silicone supply unit.
INFORMATION Silicone oil loss
To avoid leakage of silicone oil into the aspiration line
of the I/A cassette, the silicone oil collection tank must
be kept upright and replaced as soon as it is observed
that it is filled with silicone oil.
To speed up the removal of high density oil, you
can use the included Y-connector to connect two
removal groups and use them at the same time.
For more information, refer to the documents
attached to the silicone oil removal assembly.