R-evo Smart E/S/CR Installation and use manual
Cod. 1X10XXEN 2020-09-14 Rev.E
94
6. Care and maintenance
6.1 MAINTENANCE OF THE DEVICE
Replacing fuses
The fuses of the control unit are located in the mains
socket on the back of the unit. To replace the fuse,
proceed as follows:
CAUTION Risk of injury to the user!
Hot fuses could cause burns.
Before replacing the fuses, turn off the device and
allow it to cool for a few minutes.
Turn off the medical device with the power switch.
1
Pull out the power cord.
Press the lock lever (1) and remove the fuse holder.
Remove the faulty fuses and install the new ones.
INFORMATION Risk of damaging the Medical Device!
Incorrect fuses may damage the device.
Use only specified fuses! Refer to the label on the rear
panel for detailed information.
Install the fuse holder again. Lock the lever (1) making it
snap into place.
Insert the power cord.
Turn on the unit with the power switch.