the instrument to the equipotential connector of the
operating room using an equipotential bonding cable. If
necessary, have a second ground reference in the
event of a malfunction of the main earth plug.
• The power plug is used as a means to electrically isolate
all poles simultaneously from the MAINS SUPPLY. Do not
place the appliance in such a way that it is difficult to
disconnect the power plug from the wall socket
• Check the parameter settings of the selected user
profile to prevent unwanted behaviour of the Medical
Device.
After each use
• At the end of the operating session, disconnect all the
surgical accessories. Reinsert the instrument tray into
the appropriate compartment on the console.
• Use the main power switch to turn off the Medical
Device.
• The main switch must be turned off when not in use. The
power cord plug must be removed from the mains
supply.
• (CR only) Disconnect the compressed air supply tube
from the operating room system.