R-evo Smart E/S/CR Installation and use manual
Cod. 1X10XXEN 2020-09-14 Rev.E
37
the instrument to the equipotential connector of the
operating room using an equipotential bonding cable. If
necessary, have a second ground reference in the
event of a malfunction of the main earth plug.
The power plug is used as a means to electrically isolate
all poles simultaneously from the MAINS SUPPLY. Do not
place the appliance in such a way that it is difficult to
disconnect the power plug from the wall socket
Check the parameter settings of the selected user
profile to prevent unwanted behaviour of the Medical
Device.
After each use
At the end of the operating session, disconnect all the
surgical accessories. Reinsert the instrument tray into
the appropriate compartment on the console.
Use the main power switch to turn off the Medical
Device.
The main switch must be turned off when not in use. The
power cord plug must be removed from the mains
supply.
(CR only) Disconnect the compressed air supply tube
from the operating room system.