R-evo Smart E/S/CR Installation and use manual
Cod. 1X10XXEN 2020-09-14 Rev.E
11
Carefully follow the instructions when installing and
using the unit in order to prevent harmful interference
by with other devices.
This Medical Device has been tested and found
compliant to the emission and immunity limits
established by the IEC60601-1-2:2014 standard for
electro-medical devices. These limits are conceived to
ensure proper protection from interferences in a typical
medical environment. However, interferences may still
exist in a specific installation. If the system causes
harmful interference with the function of other devices
(can be detected by turning the unit off and on again),
the user is encouraged to try to remedy the
interference by one or more of the following measures:
Reorient or relocate other devices.
Increase the distance between the devices.
Connect the unit to a power outlet other
than the one to which the other devices
are connected.
Contact your dealer or the technical service of
OPTIKON 2000 S.p.A.
The sound emission capacity of the device is tested at
startup.
Verify that a beep sounds is emitted when initializing the
system.
Any malfunctions occurring during use of the medical
device are indicated by a message on the display and by
alarm signals. Resolve the malfunction and confirm the
message by pressing the corresponding key in the
display. If the malfunction can not be eliminated, or if
the error occurs again, do not continue to use the
device, but put up signs indicating that it is "out of
order" and contact your dealer or the OPTIKON 2000
S.p.A. technical service.
Do not pull on the power cables or other connecting
cables.
When moving the Medical Device make sure that no
tubes are crushed or pulled out.
Portable and mobile radio communication devices can
affect operation of the Medical Device