• Remove the spike of the infusion set from the bottle,
taking care not to touch it with your hands or any other
unsterile material.
• Insert the spike of the infusion set into the cap of the
new bottle and hook it onto the IV pole.
• If the drip chamber becomes completely empty,
squeeze it to fill it with balanced salt solution until it is
approximately 50% full.
• Re-open the infusion clamp.
• If air bubbles are observed in the irrigation tubings, ask
the surgeon to activate irrigation until the bubbles are
expelled, before reinserting the handpiece into the
patient's eye.
CONTROLLED INFUSION
• Stop the surgical procedure and remove the handpiece
from the incision.
• Close the infusion tube clamp.
• Disconnect the air tube from the "CONTROLLED" outlet
located on the R-Evo Smart pump plate.
• Turn off the near-empty infusion liquid container from
its support.
• Remove the spike of the infusion set, taking care not to
touch the needle with your hands or any other unsterile
material.
• Insert the spike of the infusion set into the cap of a new
container and hang it from the outflow nozzle holder.
• Connect the air tube from the "CONTROLLED" output of
the R-Evo Smart medical device.
• Re-open the irrigation clamp.
• If air bubbles are observed in the irrigation tubings, ask
the surgeon to activate irrigation until the bubbles are
expelled, before reinserting the handpiece into the
patient's eye.