R-evo Smart E/S/CR Installation and use manual
Cod. 1X10XXEN 2020-09-14 Rev.E
22
Remove the spike of the infusion set from the bottle,
taking care not to touch it with your hands or any other
unsterile material.
Insert the spike of the infusion set into the cap of the
new bottle and hook it onto the IV pole.
If the drip chamber becomes completely empty,
squeeze it to fill it with balanced salt solution until it is
approximately 50% full.
Re-open the infusion clamp.
If air bubbles are observed in the irrigation tubings, ask
the surgeon to activate irrigation until the bubbles are
expelled, before reinserting the handpiece into the
patient's eye.
CONTROLLED INFUSION
Stop the surgical procedure and remove the handpiece
from the incision.
Close the infusion tube clamp.
Disconnect the air tube from the "CONTROLLED" outlet
located on the R-Evo Smart pump plate.
Turn off the near-empty infusion liquid container from
its support.
Remove the spike of the infusion set, taking care not to
touch the needle with your hands or any other unsterile
material.
Insert the spike of the infusion set into the cap of a new
container and hang it from the outflow nozzle holder.
Connect the air tube from the "CONTROLLED" output of
the R-Evo Smart medical device.
Re-open the irrigation clamp.
If air bubbles are observed in the irrigation tubings, ask
the surgeon to activate irrigation until the bubbles are
expelled, before reinserting the handpiece into the
patient's eye.