R-evo Smart E/S/CR Installation and use manual
Cod. 1X10XXEN 2020-09-14 Rev.E
25
Vitrectomy applications
Never activate the vitrectomy handpiece with the blade
in the air. The vitrectome must always be tested with
the blade in a sterile solution container. The test
carried out in the air will cause irreparable damage to
the cutting system.
Illumination (R-Evo Smart CR only)
WARNING Risk of injury to the patient's eye!
Adjust the intensity of the light irradiation and the
corresponding exposure duration by selecting
appropriate irradiation settings.
WARNING Risk of injury to the patient's eye!
Damage to the retina due to excessive power.
Although the fibre optic endo-ocular illumination
system is designed not to emit infrared and ultraviolet
radiation except in a small quantity, to avoid possible
damage to the retina, always use the minimum
intensity compatible with the surgical applications.
INFORMATION Risk of minor injury to the patient!
The light output connectors are hot when the light is on.
Switch off the light and wait several minutes before
touching the light output connectors!
To reduce the risk of damage to the retina, the tip of
the fibre optic light probe should not be located in the
immediate vicinity of the retina.