Powering the device on
Before turning on the device, check the following:
• The compressed air line has been connected (R-Evo
Smart S and CR only).
• The power cord has been connected.
• The pedal has been connected.
WARNING Risk of injury to the patient's eye!
• An audible air leak at start-up may indicate a
malfunction of the silicone section.
Contact your dealer or the technical service of OPTIKON
2000 S.p.A.
CAUTION Functional test
• The software checks the validity of the calibration at
power on. If there is any discrepancy, the default safe
calibration is stored and you are prompted to calibrate
the device (5 beeps).
Contact your dealer or the technical service of OPTIKON
2000 S.p.A.
To use the Medical Device, proceed as follows:
• Set the power switch on the rear panel to the on position
(I).
• Press the power button on the front panel until the
indicator lamp lights up.
• The electric irrigation pole rises and lowers by a few
centimetres.
• The device performs a self-test, during which a start
splash screen shows the R-EVO Smart logo.
• When the device is on, the first menu appears on the
screen. User can choose among “Surgery”, “System
settings” and “Service”.
•
At the first start-up, the R-evo Smart operating
interface is set in English, the “Surgery” page shows the
standard user profiles, "Default", the user can choose a
different language in the “System settings” page.