SERVICE MANUAL
Safety aspects
Ow-RX (Rev. 1)
6
Immunity test
Test level
IEC 60601-1-2
Compliance
level
EMC environment of use
The RF portable and mobile
communications units should not be
used closer to any part of the Ow-RX,
including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance:
Radiated RF
IEC 61000-4-3
3 V/m
from 80 MHz to
2.5 GHz
3 V/m
d = 1.2 x
P from 80 MHz to
800 MHz
d = 2.3 x
P from 800 MHz to
2.5 GHz
Conducted RF
IEC 61000-4-6
3 V
from 150 kHz to
80 MHz
3 V
d = 1.2 x
P
where "P" is the maximum rated
output of the transmitter in watts (W)
according to the transmitter
manufacturer and "d" is the
recommended separation distance in
meters (m).
The field strength of the fixed RF
transmitters, determined by an on-site
electromagnetic survey, should be
lower than the compliance level in
each frequency range.
Interference may be verified near
devices marked with the following
symbol: