EasyManua.ls Logo

Owandy RX - Remote Timer Connection

Owandy RX
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SERVICE MANUAL
Installation
(Rev. 1) Ow-RX
55
6.5.2.2 Remote Timer connection
1. Assemble the X-ray button complete of its support (2 - Figure 6-21)
on the Remote Timer plate.
2. Connect the RJ 45 cable to the Remote Interconnection Timer board
(A9) – connector X9 (Figure 6-21).
3. Assemble the two pins lock (1 - Figure 6-21) on the Timer cover (1 -
Figure 6-1); these pins (code 2100511300) are included in the
mounting hardware packaging.
4. Connect the flat cable between CPU board – connector X1 – and A9
board – connector X1. Connect the X-ray button cable to the
connector X6 on the CPU board.
5. Place the cover on the remote support taking care that the cover is
positioned under the upper plate (3 - Figure 6-21). Push the cover
against the locking pins and tighten the two upper screws.
X1
X9
Remote Interconnection
Timer board (A9)
2
3
1
X1 X6
Figure 6-21

Table of Contents