4. TECHNICAL DATA
Technical characteristics
Equipment Ow-RX
Manufacturer OWANDY RADIOLOGY
77183 Croissy-Beaubourg - France
Class
Class I with type B applied parts
(according to EN 60601-1
classification)
Protection degree IPX0 standard device
Line voltage
99-264 V∼
Rated voltage
110-240 V∼
Line frequency 50 / 60 Hz
Maximum line current 5.2 A rms impulsive
@ 115 V ∼
2.5 A rms impulsive
@ 230 V ∼
Technical factors for maximum line
current
70kV, 6mA
Absorbed power 583W (584VA) 566W (570VA)
Maximum apparent line resistance
0.4 Ω (99-132 V∼) 0.8 Ω (198-264 V∼)
Mains fuse 6.3 AT
Selectable times from 0.01 to 2.00 s in 36 steps
Automatic selection 882 pre-programmed times (7 anatomic -
3 sizes - 3kV - 2 SID- 3 receptors)
Time accuracy
± 5 % ± 2 ms
High voltage values 60-65-70 kVp selectable
Tubehead current 6 mA
kV accuracy
± 8 % @ rated voltage
Tubehead anodic current accuracy
± 10 % @ rated voltage
Maximum exposure time 2.0 s
Timer size
284×253×123.3 mm