SERVICE MANUAL
Troubleshooting
(Rev. 1) Ow-RX
73
9.1.3 NON fatal errors
Situations which do not directly effect the safety of the operator, patient
or the system are considered as re-settable anomalies.
The error condition prevents additional exposures until it is reset
by pressing the "Increase" key.
Displayed
message
ANOMALY type Corrective actions
CH0
Memory checksum error
(EEPROM)
If present at the first start-up of the
equipment after replacing EEPROM, follow
the displayed procedure.
On the contrary, if present during normal
working, replace the EEPROM, if not
enough, replace CPU board.
E11
X-ray button active after
the cooling phase
Release the X-ray button and follow the
displayed procedure.
If the X-ray button is not pressed, switch
off the system and replace the button.
E12
Release of X-ray button
during emission
Reset the error and repeat exposure.
Make sure NOT to release the X-ray button
before the buzzer stops.
If the message is still present, replace the
X-ray button.
E13
Release of X-ray button
during pre-heating phase
Reset the error and repeat exposure.
Make sure NOT to release the X-ray button
before the buzzer stops.
If the message is still present, replace the
X-ray button.