SERVICE MANUAL
Set-up
Ow-RX (Rev. 1)
62
8. SET-UP
The Ow-RX system has a modifiable configuration that can be adapted to
specific usage needs.
To enter the "set-up" function press the "Increase" and
"Decrease" keys simultaneously during the software / hardware
version display phase; the display will show:
System Setup
PXX: selected parameter description
Value = selected parameter value
Press "Function" key to go to the following parameter; touch
either the "Increase" or "Decrease" keys to change the displayed
parameter value.
Press "Function" key and then confirm (Increase key) or cancel (decrease
key) the new setting.
Press and hold pressed the "Function" key to exit the setup program, the
system will restart.
P00:
Stand-by timeout (min)
This parameter indicates the time in minutes after that the system enter
in "sleep" condition.
With "Value = 0", the system is always in the "Ready for X-rays"
conditions. The selectable value changes between 0 and 5 (default 2).
NOTE:
When Value = 0, the green "Ready for X-rays" LED is always ON.