P01:
Duty cycle
This is the value of parameter K (value selectable between 0 and 80)
which, multiplied by the current value of exposure time, calculates the
value of the cool down time between consecutive exposures.
COOL DOWN BETWEEN EXPOSURES = Exposure time x K
Default value = 60 (sixty times the exposure time).
WARNING:
This parameter must not be changed from the factory configuration.
Changing on this parameter may effect on the equipment life time.
P02:
Exposure Counter
This parameter shows the number of the performed exposure. The
counter can be reset to zero by pressing the "Size Selection" and
"Tooth automatic selection" keys together; the operation requires
confirmation, by pressing the "Increase" key.
P03:
Film sensitivity
This parameter adapts the exposure times associated to Anatomic and
Size selections to the film speed used by the user.
The available values are shown in the following table:
Value
(multiplies the
pre-set exp. time)
Film type
0.75 --
1.00 F
1.25 E
1.50 --
2.00 D
Default value = 1.00 (F)