SERVICE MANUAL
Replacing parts
Ow-RX (Rev. 1)
74
10. REPLACING PARTS
NOTE:
The Service Engineer has to take special care for all what concerns
electrical safety of the device and must make sure of restoring all
provisions for electrical safety which may be affected during a service
intervention and to solicit the customer to have the electrical safety tests
repeated every time the intervention has caused the replacement of
important parts or the intervention has significantly affected safety
provisions of the device.
10.1 Replacing the tubehead and the Driver board
WARNING:
The tubehead and the Diver board are matched in the factory and
therefore MUST be replaced as a couple.
WARNING:
Before proceeding with the replacement of the tubehead, close and tie the
two sections of the scissors arm together.
Failure to follow these instruction may harm both the technician and
damage the arm itself.
NOTE:
Before carrying out the following operations, disconnect the equipment
from the line.
NOTE:
No specific actions are required after replacing the tubehead and the
Driver board.