37
Basic function
Range
Minimum
resolution
Accuracy
2.0000V
0.1mV
±(0.3%+5dig)
20.000V
1mV
200.00V
0.01V
1000.0V
0.1V
AC voltage
[1]
200.00mV
0.01mV
±(0.8%+10dig)
2.0000V
0.1mV
20.000V
1mV
200.00V
0.01V
750.0V
0.1V
±(1%+10dig)
Frequency range40Hz-1000Hz
DC current
200.00mA
0.01mA
±(0.8%+10dig)
10.000A
1mA
±(2.5%+10dig)
Overload protection:
mA function: self-healing fuse 400 mA/250 V; Ampere function: 10
A/600 V, D5.2*20, fast-acting fuse
AC current
[1]
200.00mA
0.01mA
±(1%+10dig)
10.000A
1mA
±(2.8%+10dig)
Frequency range
40Hz-1000Hz
Overload protection:
mA function: self-healing fuse 400 mA/250 V; Ampere function: 10
A/600 V, D5.2*20, fast-acting fuse
Resistance
200.00Ω
0.01Ω
±(0.8%+10dig)
2.0000kΩ
0.1Ω
±(0.8%+5dig)
20.000kΩ
±(0.8%+3dig)
200.00kΩ
10Ω
2.0000MΩ
0.1kΩ
20.000MΩ
1kΩ
±(1%+3dig)
100.00MΩ
0.01MΩ
±(5%+10dig)
Capacitance
[1]
20.000nF
1pF
±(3.0%+10dig)
200.00nF
10pF
2.0000μF
0.1nF
20.000μF
1nF
200.00μF
10nF
2.0000mF
0.1uF
Others
On/Off test
(<50Ω)