EasyManua.ls Logo

Owon XDS4000 Series - Page 114

Owon XDS4000 Series
145 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5.Use the Arbitrary Function Generator
108
Output V
CH2
(Note: The menus are not for selecting menu items,
only for showing information.)
Min Freq
Fine
Turn Multipurpose knob to set frequency sweep
min value. Check to enable fine-tuning.
Max Freq
Fine
Turn Multipurpose knob to set frequency sweep
max value. Check to enable fine-tuning.
Amplitude
Fine
Turn Multipurpose knob to set waveform
generator amplitude. Check to enable fine-tuning.
Note: The amplitude range is 2 mVpp to 6 Vpp.
Chart
Gain Scale
Adjust scale of gain plot, range from 5dB to
500dB.
Gain Offset
Adjust offset of gain plot, range from -250dB to
250dB.
Phase Scale
Adjust scale of phase plot, range from 5° to 180°.
Phase Offset
Adjust offset of phase plot, range from -180° to
180°.
Autoscale
Autoscale gain and phase plots.
Analysis
Run/stop the analysis.
F: Frequency; G: Gain; P: Phase.
To run the frequency response analysis, do the following:
1. The waveform generator output is connected to a device under test. The input
to the device is probed by CH1of the oscilloscope. The output from the device
is probed by CH2of the oscilloscope.
2. Push the Utility button, select Function in the bottom menu, select FRA in the
left menu
3. In the bottom menu, select FRA. In the right menu, check FRA, and set
other menu items.
4. In the bottom menu, select Setup or Chart. In the right menu, set the menu
items.
5. In the bottom menu, check Analysis to run the analysis.

Table of Contents