4.Advanced User Guidebook
Set idle time before the edge counting in Nth Edge
Trigger. Adjust M knob or click to set idle time
press panel button or click to
move cursor to choose which digit to be set.
Available range is 30ns-10s, default 100ns.
Set the edge number value of “N” in Nth Edge
trigger.
Auto
Normal
Single
Holdoff
Noise Reject
ON OFF
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform then
stop
100 ns - 10 s, adjust M knob or click to set
time interval before another trigger occur, press
panel button or click move
cursor to choose which digit to be set.
Turn ON/OFF Noise Reject.
(Range 0.3div to10div)
Logic Trigger
Trigger according to logic relation.
In Logic Trigger mode, the trigger setting information is displayed on bottom
right of the screen, for example,
,
indicates that trigger type is Logic, logic mode is AND, CH1 high level and
trigger level is 0.00mV.
Logic Trigger menu list:
Set vertical channel trigger type as Logic trigger.
Set logic mode as AND.
Set logic mode as OR.
Set logic mode as XNOR.
Set logic mode as XOR.