4.Advanced User Guidebook
43
Mode
Holdoff
Auto
Normal
Single
Holdoff
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform then
stop
100 ns 10 s, turn the M knob or click to set
time interval before another trigger occur, press
panel button or click to move
cursor to choose which digit to be set.
5.Runt Trigger
Trigger pulses that pass through one trigger level but fail to pass through the
other trigger level. Shown as below figure,
In Runt Trigger mode, the trigger setting information is displayed on bottom
right of the screen, for example, , indicates that trigger
type is runt, trigger source is CH1, polarity is positive, 0.00mV is the differential
between up level and low level threshold.
Runt Trigger
Runt Trigger menu list:
MENU
SETTING
INSTRUCTION
Single
Mode
Runt
Set vertical channel trigger type as runt trigger.
Source
CH1
CH2
CH3
CH4
Select CH1 as the trigger source
Select CH2 as the trigger source
Select CH3 as the trigger source
Select CH4 as the trigger source