4.Advanced User Guidebook
Auto
Normal
Single
Holdoff
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform then
stop
100 ns – 10 s, turn the M knob or click to set
time interval before another trigger occur, press
panel button or click to move
cursor to choose which digit to be set.
5.Runt Trigger
Trigger pulses that pass through one trigger level but fail to pass through the
other trigger level. Shown as below figure,
In Runt Trigger mode, the trigger setting information is displayed on bottom
right of the screen, for example, , indicates that trigger
type is runt, trigger source is CH1, polarity is positive, 0.00mV is the differential
between up level and low level threshold.
Runt Trigger
Runt Trigger menu list:
Set vertical channel trigger type as runt trigger.
Select CH1 as the trigger source
Select CH2 as the trigger source
Select CH3 as the trigger source
Select CH4 as the trigger source