4.Advanced User Guidebook
Windows Trigger menu list:
Set vertical channel trigger type as Windows
trigger.
Select CH1 as the trigger source
Select CH2 as the trigger source
Select CH3 as the trigger source
Select CH4 as the trigger source
Adjust the M knob or click to set the up level
threshold.
Adjust the M knob or click to set the low level
threshold.
Positive Polarity, the oscilloscope triggers on the
positive Windows pulse.
Negative Polarity, the oscilloscope triggers on the
negative Windows pulse.
Enter: Triggers when the trigger signal enters the
specified trigger level range.
Exit: Triggers when the trigger signal exits the
specified trigger level range.
Time: Specify the hold time of the input signal after
entering the specified trigger level. The
oscilloscope triggers when the accumulated hold
time is greater than the windows time. Available
range is 30ns-10s, default 100ns.
Auto
Normal
Single
Holdoff
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform then
stop
100 ns - 10 s, adjust M knob or click to set
time interval before another trigger occur, press
panel button or click to move
cursor to choose which digit to be set.