4.Advanced User Guidebook
indicates that trigger type is UART/RS232, CH1 trigger level is 1.80V.
Format as shown in the figure below,
UART/RS232 Trigger
UART/RS232 Trigger menu list:
Set vertical channel bus type as
UART/RS232 trigger.
Select CH1 as the trigger source.
Select CH2 as the trigger source.
Select CH3 as the trigger source.
Select CH4 as the trigger source.
Select polarity of data transmission as
Normal. Select polarity of data
transmission as Inverted.
Trigger on the start frame of position. After
choosing this condition, press Configure to
enter detailed settings.
Trigger when error frame is detected. After
choosing this condition, press Configure to
enter detailed settings.
Trigger when Chk Error is detected. After
choosing this condition, press Configure to
enter detailed settings.
Trigger on the last bit of the preset data.
After choosing this condition, press
Configure to enter detailed settings.
Common Baud: adjust M knob to choose
common baud.
Custom Baud: adjust M knob to choose
baud, ranges from 50 to 10,000,000.
Stop Bit:Select “1”or ”2”.
Parity: “NO””EVEN””ODD”
Common Baud: adjust M knob to choose
common baud.
Custom Baud: adjust M knob to choose
baud, ranges from 50 to 10,000,000.