4.Advanced User Guidebook
54
sample point, and signal rate of the CAN signal.
In CAN bus trigger mode, the trigger setting information is displayed on
bottom right of the screen, for example, , indicates
that trigger type is CAN, CH1 trigger level is -126 mV.
CAN Trigger menu list:
MENU
SETTING
INSTRUCTION
Bus
Type
CAN
Set vertical channel bus type as CAN trigger.
Input
Source
CH1
CH2
CH3
CH4
Select CH1 as the trigger source.
Select CH2 as the trigger source.
Select CH3 as the trigger source.
Select CH4 as the trigger source.
Type
CAN_H
CAN_L
TX
RX
Actual CAN_H bus signal.
Actual CAN_L bus signal.
Transmission signal on the CAN signal
line.
Received signal on the CAN signal line.
Sample
Point
Turn the M knob (or tap on in touchscreen) to
set the Sample point, which is a point within a bit’s
time. The oscilloscope samples the bit level at this
point. “Sample point” is represented by the
percentage of “the time from the start of the bit’s time
to the sample point time” in the “bit’s time”. The
range is 5% to 95%.
Common
Baud
Turn the M knob to select from the Baud list on the
left.
Custom
Baud
Turn the M knob (or tap on in touchscreen) to
set the Baud. The range is 10,000 to 1,000,000.
Tip: You can select the nearest value in Common
Baud, and then adjust it in this menu.
Conditio
n
Start
Trigger on the start frame of the data frame.
Type
Type
(Bottom
menu)
Data
Trigger on the selected
frame.
Remote
Error
Overloa
d