4.Advanced User Guidebook
parameters based on the characteristics of the signal. When the
parameters are set correctly, the information carried by the signal will be
displayed.
Tip: If there are repetitive menu items in both trigger menu and decoding menu,
you can set anyone of them, the other will be changed synchronously.
Note:
⚫ Use the Trigger Level knob to adjust the thresholds of bus trigger and bus
decoding.
⚫ When decoding, if "Parity" is not set to "None", and the check bit error is
detected, two red error marks will be displayed in the corresponding
position in the waveform.
UART/RS232 Decoding menu list:
Set bus type of decoding as UART/RS232.
Turn the M knob to select from the Baud list on
the left.
Turn the M knob (or tap on in touchscreen) to
set the Baud. The range is 50 to 10,000,000.
Tip: You can select the nearest value in Common
Baud, and then adjust it in this menu.
Set the data width of each frame to match the
signal. It can be set to 5, 6, 7 or 8.
set the even-odd check mode to match the
polarity used by the signal.
Set the display format of the bus.
Select "ON" to display the event table.
If a USB storage device is currently connected to
the instrument, save the event table data in a .csv
(spreadsheet) formatted file on the external USB
storage device.
Select "ON" to display the ASCII table.