Select the types need to be deleted.
The selected type and source are shown
in the Remove menu on the right.
Measure
Only if the waveform channel is in the ON state, the measurement can be
performed. The automatic measurement can not be performed in the
following situation: 1) On the saved waveform. 2) On Waveform Math
waveform. 3) On the Video trigger mode.
On the Scan format, period and frequency can not be measured.
Measure the period, the frequency of the CH1, following the steps below:
1. Push the Measure front panel button to show the Measure menu.
2. Select Add in the bottom menu.
3. In the left Type menu, turn the M knob to select Period.
4. In the right menu, select CH1 in the Source menu item.
5. In the right menu, select Add. The period type is added.
6. In the left Type menu, turn the M knob to select Frequency.
7. In the right menu, select CH1 in the Source menu item.
8. In the right menu, select Add. The frequency type is added.
The measured value will be displayed at the bottom left of the screen